Contact an Account ManagerHydrocarbons in Water acc. to ISO DIS 9377-4 / DIN H-53
Summary
Type: SPE
Application #: 302090
Catalog#: 730249G - SPE Cartridges, Chromabond, Na2SO4/Florisil, 6ml, Glass, 2000/2000mg 30/pk
Detection: Further analysis: According to our GC application 210600 on OPTIMA 1, 25 m, ID 0.32 mm, 0.25 µm
Conditions: Internal standard solution: Solve 20 mg n-tetracontane (C40H82) in cyclohexane, add 20 ml n-nonane (C9H20) and fill up to 1 litre with cyclohexane. To prepare the extraction solution dilute the standa
Authors: SPE department
Source: Macherey-Nagel
Details
Legend: